Portable image-based detection system for Upconverting phosphor immune lateral flow assay

Konferenz: BIBE 2019 - The Third International Conference on Biological Information and Biomedical Engineering
20.06.2019 - 22.06.2019 in Hangzhou, China

Tagungsband: BIBE 2019

Seiten: 7Sprache: EnglischTyp: PDF

Persönliche VDE-Mitglieder erhalten auf diesen Artikel 10% Rabatt

Autoren:
Li, Chao; Chen, Meng; Cheng, Zhi; Chen, Feng; Wu, Taihu; Du, Yaohua (Institute of Medical Equipment, Academy of Military Medical Sciences, Tianjin, China)

Inhalt:
A portable image-based detection system for Up-converting phosphor immune lateral flow assay is introduced. A commecial low light level CMOS sensor is chosen as the detector, which is low cost. The excitation light source is a 980nm and 500mw laser diode. In order to minish the error induced by the un-uniformity of laser irradiation, a diffuser is inserted into the beam and a calibration algorithm is developed. The residual error from the un-uniformity irradiation is less than 1%. Benefit from the adjustable of the exposure time, the integrated dynamic range of the system is as high as 106dB. The repeat test error for the very low signal is 1% (variation coefficient). And the linear correlation coefficient between the tested T/C value and sample concentration is 0.998. Compared to the traditional instrument, the detection system is static, low cost, portable and high-precision.