A Model-Based Synthetic Aperture Focusing Technique for High- Frequency Ultrasound Imaging
Konferenz: EUSAR 2010 - 8th European Conference on Synthetic Aperture Radar
07.06.2010 - 10.06.2010 in Aachen, Germany
Tagungsband: EUSAR 2010
Seiten: 4Sprache: EnglischTyp: PDF
Persönliche VDE-Mitglieder erhalten auf diesen Artikel 10% Rabatt
Autoren:
Opretzka, Jörn; Vogt, Michael; Ermert, Helmut (High Frequency Engineering Research Group, Ruhr-Universität Bochum, Germany)
Inhalt:
In this contribution a mono-static synthetic aperture focusing image reconstruction technique for high-frequency ultrasound imaging systems with fixed-focus single element transducers is suggested. It is based on the correlation of recorded echo signals with the simulated point spread function of the system. The simulation is based on the real transducer properties and takes the echo signal's phase and direction-dependent magnitude into account for reconstruction. The proposed concept has been evaluated by measurements with a 20 MHz ultrasound system on a wire phantom and with small animal imaging experiments. By synthetic aperture focusing the lateral resolution is improved and nearly constant over the whole depth range.