Cover DIN EN 60601-2-54 VDE 0750-2-54 Berichtigung 1:2010-07
larger

DIN EN 60601-2-54 VDE 0750-2-54 Berichtigung 1:2010-07

Corrigendum to DIN EN 60601-2-54 (VDE 0750-2-54):2010-05; (IEC-Cor.:2010 to IEC 60601-2-54:2009)

Class/Status: Standard, withdrawn
Released: 2010-07
VDE Art. No.: 0750165

Due to IEC Corrigendum 1:2010 DIN EN 60601-2-54 (VDE 0750-2-54):2010-05 has to be amended by the following sentence in 203.11.102 b): use a phantom having an attenuation equivalent of 40 mm AI, positioned in the X-ray beam as close as possible to the focal spot.

Standard
withdrawn:

DIN EN 60601-2-43 VDE 0750-2-43:2011-03

Medical electrical equipment

Part 2-43: Particular requirements for basic safety and essential performance of X-ray equipment for interventional procedures

119.08 € 
Add Document
Standard
withdrawn:

DIN EN 60601-2-54 VDE 0750-2-54:2016-07

Medical electrical equipment

Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy

149.73 € 